FEATURE: |
|
Three channel Interpretive ECG machine |
Simultaneously 12 leads acquisition |
8 Patients save/copy/communication |
Multiple operation mode (Manual/Auto/Ref.lead) |
Large LCD to display 12 leads ECG waveform |
Built-in standard RS232 Interface (USB is optional) |
DC/AC power with built-in rechargeable lithium battery |
|
|
SPECIFICATION: |
|
Lead |
Standard 12 leads |
Acquisition |
Simultaneous 12 leads |
Resolution |
12bit/1000Hz |
Operation mode |
Manual mode , 3 Auto Modes (with interpretive report) |
Filter |
AC , EMG Filter |
Drift filter |
Anti-drift system |
CMMR |
> 60dB |
|
> 100dB(with AC filter) |
Input circuit |
Floating Protection circuit against Defibrillator effect |
Input circuit current |
< 0.1µA |
Input Impedance |
> 50 M¦¸ |
Patient current leakage |
< 10µA |
Calibrating Voltage |
1mV +/- 3% |
Time constant |
> 3.2 s |
Frequency response |
0.05Hz~150Hz |
Noise level |
< 15µVp-p |
Threshold |
< 20µV |
Paper speed |
25, 50mm/s (+/- 3%) |
Sensitivity |
Auto , 2.5,5,10,20mm/mV |
Recorder |
High resolution thermal array printer |
Paper specification |
63mm x 30m roll |
LCD display |
Display menu, operation status and ECG waveform |
Safety level |
GB9706.1-1995 class I , type CF |
Power supply |
AC 220V/110V +/- 10% 50/60Hz +/- 1Hz 30VA |
|
DC 14.4V-1200mAh rechargeable lithium battery |
Dimension |
288 x 209 x 62mm |
Dimension (carton) |
380 x 300 x 210mm |
Net Weight |
3kgs |
Weight |
4.5kgs |
|
|
|
|
Standard Accessories: |
|
Patient Cable, Thermal Printing Paper, Limb Electrodes, Chest Electrodes, Power Supply Cable, |
Grounding Wire |
|
FEATURE: |
|
Single/Three channel digital ECG machine with measurement |
Simultaneously 12 leads acquisition |
Multiple operation mode (Manual/Auto/Ref.lead) |
LCD to display operation status and ECG waveform |
Built-in standard RS232 Interface (USB is optional) |
DC/AC power with built-in rechargeable lithium battery |
|
|
SPECIFICATION: |
|
Lead |
Standard 12 leads |
Acquisition |
Simultaneous 12 leads |
Resolution |
12bit/1000Hz |
Operation mode |
Manual mode , 3 Auto Modes (with measurement report ) |
Filter |
AC , EMG Filter |
Drift filter |
Anti-drift system |
CMMR |
> 60dB |
|
> 100dB(with AC filter) |
Input circuit |
Floating Protection circuit against Defibrillator effect |
Input circuit current |
< 0.1µA |
Input Impedance |
> 50 M¦¸ |
Patient current leakage |
< 10µA |
Calibrating Voltage |
1mV +/- 3% |
Time constant |
> 3.2 s |
Frequency response |
0.05Hz~150Hz |
Noise level |
< 15µVp-p |
Threshold |
< 20µV |
Paper speed |
25, 50mm/s (+/- 3%) |
Sensitivity |
Auto , 2.5,5,10,20mm/mV |
Recorder |
High resolution thermal array printer |
Paper specification |
63mm x 30m roll |
LCD display |
Display menu, operation status and ECG waveform |
Safety level |
GB9706.1-1995 class I , type CF |
Power supply |
AC 220V/110V +/- 10% 50/60Hz +/- 1Hz 30VA |
|
DC 14.4V-1200mAh rechargeable lithium battery |
Dimension |
288 x 209 x 62mm |
Dimension (carton) |
380 x 300 x 210mm |
Net Weight |
3kgs |
Weight |
4.5kgs |
|
|
|
|
Standard Accessories: |
|
Patient Cable, Thermal Printing Paper, Limb Electrodes, Chest Electrodes, Power Supply Cable, |
Grounding Wire |
|